Nihon Kohden · Resuscitation and AEDs
cardiolife TEC-1021
TEC-1021
The version without pacer of Nihon Kohden’s cardiolife TEC-1000 series: monitor-defibrillator for the hospital, 2 to 270 J, 10.4-inch touchscreen.
- Energy
- 2 to 270 J
- Screen
- 10.4-inch colour screen
- Battery
- 300 minutes of monitoring
- Price
- On quote
For owners of this device
Do you have this device?
Choose what you need. We fill in cardiolife TEC-1021 for you.
Defibrillator without pacer
The cardiolife TEC-1021 is the version without pacer of the cardiolife TEC-1000 series from Nihon Kohden, a monitor-defibrillator for hospitals. The TEC-1031 also has transcutaneous pacing.
The device ends ventricular fibrillation and pulseless ventricular tachycardia with a short, powerful shock. It also measures the ECG and can perform asynchronous defibrillation and synchronised cardioversion, for atrial flutter among other things.
What the device can do
- Energy: selectable from 2 to 270 J, with fast charging to 150 J in under 3 seconds (full battery) and ECG recovery within 3 seconds.
- AED mode: continuous VF/VT analysis in the background. For a child aged 6 or younger, use child mode.
- Operation: 10.4-inch touchscreen and a rotary knob.
- Battery: 300 minutes of continuous monitoring, or 150 shocks of 270 J.
According to the series brochure there are options for SpO₂, CO₂ and NIBP, for esCCO (estimated cardiac output) and synECi18 (18-lead ECG), and for CPR feedback with the CPR assist. All details, options and accessories are on the page cardiolife TEC-1000.
Safety and use
The device is intended for medical personnel trained in resuscitation. Do not use it in an MR examination room, with flammable anaesthetic gas or a high oxygen concentration, or in a hyperbaric oxygen chamber. Always read the manufacturer’s safety information and operator’s manual.
Which version and options suit your situation, we discuss with you beforehand.
Specifications
- Model
- cardiolife TEC-1021 from Nihon Kohden, the version without pacer of the TEC-1000 series.
- Version
- Without transcutaneous pacing. The TEC-1031 has it.
- Application
- Asynchronous defibrillation and synchronised cardioversion (for atrial flutter, among other things); ECG measurement.
- Screen
- Anti-glare 10.4-inch colour LCD with touchscreen.
- Dimensions (series)
- 333 × 289 × 225 mm (w × h × d).
- Weight (series)
- 6.4 kg, including battery and external paddles.
- Energy
- 2 to 270 J.
- Charging
- Under 3 seconds to 150 J with a full battery.
- ECG recovery
- Within 3 seconds after the defibrillation.
- AED mode
- Continuous VF/VT analysis in the background. For a child aged 6 or younger, use child mode.
- Battery
- 300 minutes of continuous monitoring or 150 shocks of 270 J. The manufacturer advises replacing the battery every 2 years.
- Options
- See the page cardiolife TEC-1000: SpO₂, CO₂ and NIBP, esCCO, synECi18, CPR feedback, data and accessories.
- Safety
- Only for medical personnel trained in resuscitation. Do not use in an MR examination room, with flammable anaesthetic gas or a high oxygen concentration, or in a hyperbaric oxygen chamber.
- Other specifications
- On request. Ask us for the full technical data of this model.
- Source
- Safety and Performance Information TEC-1021/TEC-1031 (1st edition, 20 June 2025) and brochure for the cardiolife TEC-1000 series, both from Nihon Kohden. Values marked “(series)” apply to the whole series.
Documents
Need the datasheet, instructions for use or CE declaration of cardiolife TEC-1021? We send them on request.
Request documents →About this system
One partner, from delivery to maintenance
We supply, install and maintain the device, and we supply the consumables. That gives you one point of contact for everything around your system.
Installation and instruction
On site, with your team.
Training
The content is stated in the quote.
Inspection and maintenance
We record the arrangements in your quote or service contract.
Remote support
Where your hospital allows VPN or TeamViewer, we can look in remotely.
Knowledge for this system
Other Nihon Kohden systems
Try first
Try cardiolife TEC-1021 first?
One of our colleagues will schedule a demo at your department.