Nihon Kohden · Cardiology

Cardiofax M

ECG-3350

ECG-3350 electrocardiograph with 8.0-inch colour screen, ECAPS 12C analysis, 500 samples/s and more than 180 minutes of battery life.

Screen
8.0-inch colour TFT, 800 × 480
Sampling rate
500 samples/s
Battery life
more than 180 minutes
Price
On quote
Cardiofax M

The Cardiofax M (ECG-3350) is an electrocardiograph with an 8.0-inch colour screen and the ECAPS 12C analysis program. Waveform processing works at 500 samples/s and, with the optional software QP-230E, the unit offers a synthesised 18-lead ECG (synECi18). Data can be output via LAN, USB or SD card in DICOM or PDF format.

Specifications

Model and order code
ECG-3350; order code ECG-3350-00K4
Leads
12 leads; 1 input channel
Sampling rate
Waveform processing: 500 samples/s, 1.25 μV/LSB. ECG input: 16,000 samples/s per channel. A/D conversion: 24 bit.
Filters
EMG filter 25 / 35 Hz; high-cut filter 75, 100, 150 Hz (−3 dB); AC line filter 50 / 60 Hz; baseline suppression weak 0.1 Hz (−20 dB), strong 0.1 Hz (−34 dB); time constant ≥3.2 s
Sensitivity and pacemaker detection
Sensitivity 5, 10, 20 mm/mV; pacemaker detection available
ECG input
Input impedance ≥50 MΩ at 0.67 Hz; common mode rejection ratio >105 dB; frequency characteristic 0.05 to 150 Hz (+0.4 dB / −3.0 dB); defibrillation-proof input protection
Heart rate measurement
Measurement and display range 30 to 300 bpm; accuracy ±10% (30 to 240 bpm)
Screen
8.0-inch, colour TFT LCD, 800 × 480 dots; shows 12-lead ECG, patient data, recording settings, heart rate, QRS sync marker, error messages, electrode detachment and noise
Recorder and paper
Thermal line recording, 200 dpi; channels 3, 3+Rhythm, 6, 12; speed 5, 10, 12.5, 25, 50 mm/s; paper 210 mm wide, 30 m long, Z-fold / 216 mm wide; extended recording and grid recording available
ECG analysis program
ECAPS 12C: approx. 200 findings, 5 assessment items; age groups neonate, child, adult. Including interpretation of Brugada-type ECG.
synECi18
Function for synthesised 18-lead ECG: 6 additional leads (V3R, V4R, V5R, V7, V8, V9) derived from the standard 12-lead ECG. Optional program QP-230E.
Memory
The most recent 3 minutes of all leads are stored internally; a portion can be selected for recording and analysis.
Data output and storage
Via wired LAN or built-in wireless LAN module; USB memory or SD card as external storage. Output files: DICOM, PDF (with SD card/USB memory). External printer available. Patient data can be received from an external server.
Connections
USB 1.1 type A: 2; SD card slot: 1; LAN port: 1
Power
100 to 240 V AC ±10%, 50 / 60 Hz ±2%, ≤150 VA
Battery
Battery pack 12 V; operates for more than 180 minutes; charging time within 3 hours
Dimensions and weight
255 (W) × 420 (D) × 140 (H) mm including handle; 4.7 kg
Protection against water
IPX0 (not protected)
Operating environment
Temperature 5 to 40°C; relative humidity 25 to 95% (non-condensing); air pressure 700 to 1,060 hPa

Documents

Need the datasheet, instructions for use or CE declaration of Cardiofax M? We send them on request.

Request documents →

IT and network

For your ICT department: what we know from the manufacturer's documentation. We agree the requirements for your hospital with your ICT.

Network and Wi-Fi
Via wired LAN or built-in wireless LAN module.
Interfaces (HL7, PACS, ADT)
Output files: DICOM, PDF (with SD card/USB memory).
Patient data can be received from an external server.
Other
Connections: USB 1.1 type A 2; SD card slot 1; LAN port 1.

A full IT checklist for your hospital ICT? Ask us.

Request the IT checklist →

About this system

One partner, from delivery to maintenance

We supply, install and maintain the device, and we supply the consumables. That gives you one point of contact for everything around your system.

Installation and instruction

On site, with your team.

Training

The content is stated in the quote.

Inspection and maintenance

We record the arrangements in your quote or service contract.

Remote support

Where your hospital allows VPN or TeamViewer, we can look in remotely.

Knowledge for this system

Het 12-afleidingen-ECG: zo plaatst u de elektrodenWaar horen de tien elektroden van een 12-afleidingen-ECG? Zo plaatst u de ECG-elektroden op ledematen en borstkas, herkent u een verkeerde plaatsing…Read the article
Holter- en inspannings-ECG: wanneer een rust-ECG tekortschietWanneer zijn een Holter- en inspannings-ECG nodig? Zo verschillen beide onderzoeken van het rust-ECG, wat ze kunnen laten zien en waar u…Read the article
Wat is een ECG en wat laat het zien?Wat is een ECG, hoe werkt een 12-afleidingen-ECG en wat laat het wel en niet zien? Een uitleg voor zorgprofessionals, met de…Read the article

Other Nihon Kohden systems

Cardiofax SECG-3250 electrocardiograph with 8-inch colour screen, ECAPS 12C analysis, wireless and wired LAN output and DICOM communication.View Cardiofax S
Cardiofax CCompact ECG-3150 electrocardiograph with 5-inch colour screen, ECAPS 12C analysis and 180 minutes of battery life.View Cardiofax C
Cardiofax VElectrocardiograph for resting and exercise ECG with 12.1" touchscreen, ECAPS 12C analysis and synECi18; internal storage for 18,000 ECG files.View Cardiofax V
Not sure which system fits?One of our colleagues will help you choose, also if you have several departments.Contact us

Try first

Try Cardiofax M first?